10-48
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
Table 10.1
Instrument Status Conditions (Continued)
Clearing fault/SB
– – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – –
The [CLEAR FAULT] key was pressed after an operator-correctable
fault was detected.
Resume operation when READY is displayed in the status box and the
word Ready on the status indicator panel is illuminated in green.
__________________________________________________________
Limits were changed to correct out of range values/BL
– – – – –– – – – – – – – – Explanation/Action– – – – – – – – – – – – – – –
A mathematically incorrect limit was manually entered (using [MEANS/
LIMITS]) during setup of a QC file. The numbers entered generated a
range containing a number greater than the largest number allowed or
less than zero. Therefore, the limits were automatically changed.
Check to be sure the entered values are correct. If appropriate, recalculate
the mean and limits and enter correct values.
__________________________________________________________
Entries making upper limit = lower limit were rejected/BL
– – – – –– – – – – – – – – – Explanation/Action– – – – – – – – – – – – – –
A mathematically incorrect limit was entered (using [RANGE
ENTRY]) during setup of a QC file. The currently entered numbers are
not accepted and the previously entered numbers for the parameter(s)
remain.
Check to be sure the entered values are correct.
__________________________________________________________
Limits were exchanged to make upper > lower/BL
– – – – –– – – – – – – – – – Explanation/Action– – – – – – – – – – – – – –
A mathematically incorrect limit was entered (using [RANGE
ENTRY]) during setup of a QC file. The numbers entered caused the
upper limit to be less than the lower limit. Therefore, the numbers were
automatically exchanged.
Check to be sure the entered values are correct. If appropriate, enter
correct values.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......