CELL-DYN
3000 System Operator’s Manual
5-93
9140240E — May 1995
Chapter 5 Operating Instructions
5.
Type the appropriate sample identification information in the
<SPECIMEN ID> and <SPECIMEN NAME> (if selected) entry
fields on the Work List. Press the ENTER key on the keyboard to
save the entry in each field and advance the cursor.
6.
If selected, type the appropriate information in the <LIMIT> and
<PARAM> entry fields. Press the ENTER key on the keyboard to
save the entry in each field and advance the cursor.
NOTE: If no entries are made in these fields, the default
selections are used.
7.
After the information for all samples has been entered, place them
in the Sample Loader racks. Samples may be placed in the racks in
any order since the bar code label is used to identify them.
NOTE: The last group of samples should be placed in an end
rack so the Sample Loader will stop when all the samples have
been processed.
8.
Install the Sample Loader safety cover.
9.
Press [RETURN] to return to the
10.
Press [SPECIMEN TYPE] followed by [PATIENT].
11.
If necessary, press [CHANGE SAMPLER] to select the Closed
Mode.
12.
Press the START key on the Sample Loader.
NOTE: Samples may be run with the
displayed. As the samples are processed, the status of each
sample is displayed in the <STATUS> field on the Work List.
Additional entries may be made to the Work List while
processing takes place.
13.
The samples are automatically processed in the order in which they
were placed in the racks. The Sample Loader automatically stops
when processing is finished if the last samples were placed in an
end rack.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......