6-84
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
the <COMMENTS> line that instrument calibration was
confirmed and no changes were required. Press the Enter key on
the keyboard to save the entry and advance the cursor.
Be certain to retain all necessary documentation in the instrument
logbook.
2.
Calibration Limit
If the % difference for a particular parameter is greater than the
following values, there may be a computation error or instrument
problem as the value has exceeded the Calibration Limit.
Results GREATER THAN CALIBRATION LIMIT:
WBC
>10%
RBC
>10%
HGB
>10%
MCV
>10%
PLT
>20%
Recheck the numbers entered on the worksheets and then recheck
all calculations.
Check to see if any component was changed which could affect the
calibration as this could cause the % difference to exceed the above
limits. For example:
Shear valve
RBC/PLT transducer assembly
WBC flow cell
RBC/PLT aperture
Syringe(s)
Hemoglobin flow cell
If a component was changed, calibrate the instrument as needed
according to the directions in the following sections. If no
components have been changed, the calculations are correct and the
% difference exceeds these limits, DO NOT CALIBRATE.
CONFIRM THAT ALL PRE-CALIBRATION PROCEDURES
WERE COMPLETED AND THEN CALL THE CUSTOMER
SUPPORT CENTER FOR ASSISTANCE AT
1 (800) CELL DYN.
3.
Calibration Range
If the % difference falls within the following Calibration Range,
calibration is required for that parameter. CALIBRATE THE
INSTRUMENT as directed in the following sections, either for all
parameters or specific parameters.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......