Calibration
Chapter 6
New calibration factors are computed for each parameter by comparing
each mean to the reference value entered for that sample. If more than
one sample is used, factors are computed for each parameter from each
of the samples. All of these factors are averaged to obtain the final
calibration factor for a given parameter.
The program computes the percent difference between the existing
calibration factor and the new one it computed. The operator decides
whether to accept the new factors based on criteria derived from CELL-
DYN 3000 performance specifications.
Calibration Requirements for Auto-Cal
The Open mode should be calibrated with the CELL-DYN Calibrator.
The Closed mode is then calibrated to correlate with the Open mode
using fresh, normal whole blood samples. In order to achieve an accurate
calibration, follow the appropriate requirements as listed below:
•
Calibrator Calibration
The Calibrator should be cycled for a minimum of 5 consecutive
runs and a maximum of 10 consecutive runs in the Open mode.
•
Whole Blood Calibration
At least five different, fresh, normal Whole Blood Samples should
be used. Each sample must be cycled for a minimum of three
consecutive runs. Whole blood samples may be run in the Open or
Closed mode.
NOTE: Refer to the
Whole Blood Calibration Guidelines
section of this chapter for detailed instructions for selecting and
processing whole blood samples.
•
Closed Mode Calibration
Ten normal whole blood samples should be used. After the Open
mode has been calibrated, each sample is cycled once in the Open
mode. The results from these samples are then used to calibrate the
Closed mode.
NOTE: Complete directions for calibration of the Closed mode
are given in the
Mode to Mode Calibration Overview
procedures sections of this chapter.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......