6-82
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
10.
Press [RETURN] followed by [MAIN] to return to the
Determining the Closed Mode Mean
1.
, press [RUN].
2.
Press [CHANGE SAMPLER] to select the Closed mode.
3.
Press [SPECIMEN TYPE].
4.
Use the arrow keys on the keyboard to move the cursor to an empty
QC file and type “CLOSED MEAN” in the <FILENAME> field.
5.
Press the Enter key on the keyboard to save the name. Use the Up
Arrow key on the keyboard to move the cursor back to the
“CLOSED MEAN” file.
6.
Press [QC SPECIMEN] to return to the
7.
Run each sample one time into the “CLOSED MEAN” QC file.
8.
Press [MAIN] followed by [QC LOGS]. With the cursor at
"CLOSED MEAN" file, press [VIEW QC LOG].
9.
Press [PRINT QC LOG] to obtain a printout of the data.
10.
Press [RETURN] followed by [MAIN] to return to the
Difference Calculation
1.
Use the mean value for each parameter from the “OPEN MEAN”
and “CLOSED MEAN” file printouts for the calculations.
2.
Enter the information from step 1 on the Manual Mode to Mode
Calibration worksheet and calculate the Mode to Mode Calibration
bias for each parameter as follows:
Closed Mode Mean - Open Mode Mean
x 100 = % Difference
Open Mode Mean
3.
Enter the % difference for each parameter in the Mode to Mode
Calibration Criteria chart on the worksheet. (This worksheet may
be duplicated as needed.) The
NOTE: Delete the sign of the factor % difference before
entering it on the chart.
4.
To determine which parameters require calibration, continue with
the next
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......