3-32
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Principles of Operation
Chapter 3
Action:
Review a stained smear and follow your laboratory's review
criteria. When bands are present, they are included in the total
neutrophil count.
IG (Immature Granulocyte) — displayed next to the NEU result
Cause:
The count in the region of scatter (on the 0
°
/10
°
plot) where
immature granulocytes are typically located is >3% of the total
WBC count.
Action:
Review a stained smear and follow your laboratory's review
criteria. When IGs are present, they are included in the total
neutrophil count.
BLAST — displayed next to the LYM result
Cause:
1.
The count in the region of scatter (on the 90
°
/0
°
plot) where blasts
are typically located is >1% of the total WBC count.
2.
The MONO % is >20% of the total WBC count.
3.
The MONO % is >3% of the total WBC count, and the standard
deviation of the monocytes on the 0
°
axis exceeds expected criteria.
Action:
Review a stained smear and follow your laboratory's review
criteria. When blasts are present, they are included in the
monocyte count.
VAR LYM — displayed next to the LYM result
Cause:
The distribution of the lymphocyte cluster on the size/
complexity (0
°
/10
°
) scatterplot exceeds expected criteria.
Action:
Review a stained smear and follow your laboratory's review
criteria. When variant lymphocytes are present, they are
included in the lymphocyte count.
NOTE: This flag may be displayed singly or in combination
with the blast flag. If the flag is displayed with the blast flag, it is
displayed as VLYM/BLAST.
NRBC — displayed next to the MONO result
Cause:
The count in a predefined region below the lymphocyte cluster
on the size/complexity (0
°
/10
°
) scatterplot is >2.9% of the total
WBC count.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......