6-36
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
10.
Consecutively run the Calibrator for the remaining number of runs.
Each subsequent run must pass the Reference Check and an
additional internal Precision Check. If a run fails either check, the
result for that parameter is highlighted and no calibration factor
will be calculated for the parameter. Calibration must be repeated
for the highlighted parameter. Calibration factors will be calculated
for the remaining parameters.
NOTE: Five runs are displayed on the screen. The screen
automatically scrolls when a sixth run is completed. Press the
Left Arrow key on the keyboard to view the previous runs if
more than 5 runs were chosen in the reference screen. Press the
Right Arrow key on the keyboard to view the last five runs.
Calibration Factor Calculation
As each run of the Calibrator is completed, a Calibration Factor is
calculated and updated. This factor is displayed in the column labeled
<SPEC (N) FACTOR>.
The column labeled <MEAN FACTOR (N)> contains the average of the
calibration factors for each parameter for all of the calibrators used in the
calibration. The number after mean factor indicates the number of
Calibrators used to compute that mean factor.
NOTE: If only one calibrator is used, the specimen factor and
the mean factor are the same.
The <FACTOR % DIFF> column displays the difference between the
calibration factor currently in the instrument and the mean factor
computed by the Auto-Cal program.
When all runs of the calibrator have been completed, the message:
NEW CAL FACTOR COMPUTED. READY FOR ACCEPTANCE
is displayed on the bulletin line.
1.
Press [PRINT] to obtain a printout of the <SPEC> and <MEAN
FACTORS> and the <FACTOR % DIFF>.
2.
Enter the factor % difference for each parameter in the Calibration
Criteria Worksheet provided at the end of this procedure. (This
worksheet may be duplicated as needed.) The Calibration Criteria
are shown in
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......