5-76
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Operating Instructions Chapter 5
13.
If the results are unacceptable, repeat the run. If the results are still
unacceptable, obtain a new tube of control, be sure that it is
warmed and mixed properly according to the manufacturer’s
recommendations and again repeat the run. If the results are still
unacceptable, run the other levels of control material. If the results
on all levels are unacceptable, troubleshoot accordingly as directed
in the
14.
When the control results are acceptable, patient samples may be
analyzed.
QC (without Q Labels) Procedure
1.
, press [SPECIMEN TYPE].
2.
Use the Arrow keys on the keyboard to move the cursor to the
appropriate QC file and press [QC SPECIMEN].
3.
If necessary, press [CHANGE SAMPLER] to select the Closed
mode.
4.
Place the controls in the Sample Loader end rack in the order in
which they are to be run.
5.
Be sure that all 10 racks (5 on each side) and the safety cover are in
place and press the Start key on the Sample Loader.
NOTE: If all QC data is going to the same file, steps 6–10 may
be skipped.
6.
After the first control is aspirated, press the Pause key on the
Sample Loader.
7.
After the results are displayed, press [SPECIMEN TYPE].
8.
Use the Arrow keys on the keyboard to move the cursor to the file
for the next control to be run and press [QC SPECIMEN].
9.
Press the Start key on the Sample Loader.
10.
Repeat steps 6–9 until all levels of controls have been run.
11.
When all of the controls have been run, press [MAIN] followed by
[QC LOGS].
12.
Use the Arrow keys on the keyboard to move the cursor to the
desired file.
13.
Press [VIEW QC LOG] to display the QC log.
14.
Verify that the results are acceptable.
NOTE: Out of control results are displayed in color.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......