CELL-DYN
3000 System Operator’s Manual
9-37
9140240E — May 1995
Chapter 9
Maintenance
5.
Pull the red release lever located in front of the transducer out and to
the right to release the aperture plate. (See Figure 9.14.)
Figure 9.14:
Transducer Assembly and Aperture Plate
6.
Pull the aperture plate straight out from between the transducer
chambers to remove it from the assembly. Note the orientation of
the plate, as it must be replaced correctly. (See Figure 9.14.)
7.
Place the plate into the container of freshly prepared cleaning
solution. Submerge the aperture completely in the solution. (The
opening is located inside the red jewel embedded in center of the
plate.) Rotate the aperture brush in the opening to ensure the cleaning
solution penetrates it completely. Allow the plate to soak for five
minutes. DO NOT soak the plate longer than five minutes as
prolonged soaking may damage the aperture plate.
CAUTION: DO NOT use anything but the aperture brush
provided in the accessory kit to clean the aperture. Using other
implements may damage the aperture.
NOTE: If desired, the container (with the aperture plate and
cleaning solution) may be placed in a sonic cleaner for two to three
minutes instead of cleaning the aperture with a brush.
CAUTION: DO NOT sonicate the aperture longer than three
minutes as prolonged sonication may loosen the aperture jewel.
Mixing
Chamber
Aperture
Plate
Counting
Chamber
Release
Lever
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......