6-70
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
Mode To Mode Auto-Cal, Closed Sampler Procedure
Procedure
The calibration procedure has been divided into subsections that consist
of a series of easy-to-follow steps. Always follow the entire procedure
unless specifically directed to skip to another section.
Confirm the calibration of the Open mode before performing this
procedure.
NOTE: Be sure to complete the
sections before beginning this procedure.
Determining Reference Values
The samples used for the calibration should be run in the Open Mode as
patient samples in order to properly identify them. (Single replications of
each of 10 whole blood samples are required.) A printout of the Data Log
for those sequence numbers can be used to enter the Reference Values
and may be retained for documentation.
1.
, press [RUN].
2.
If necessary, press [CHANGE SAMPLER] to select the Open
mode.
3.
Press [SPECIMEN TYPE] then press [PATIENT].
4.
Type the sample identification for the first sample in the <NEXT
ID> field. Press the Enter key on the keyboard to save the entry and
advance the cursor. Run the sample, noting the sequence number.
5.
Repeat step 4 for each sample. Note the sequence number of the
last sample.
6.
Access the Data Log. Press [PRINT DATA LOG].
7.
Type the range of sequence numbers for the calibration samples in
the indicated fields to obtain the printout.
Starting Auto-Cal
1.
, press [RUN] then press [WORK
LIST]. Ensure that WORK LIST and BAR CODE, as indicated
in upper left corner, are both OFF.
2.
, press [CALIBRATION].
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......