5-74
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Operating Instructions Chapter 5
Sample Loader Operating Tips
1.
All Sample Loader tubing must be connected before turning on the
Sample Loader, initializing it, or changing modes.
2.
All samples must be properly mixed before they are placed in the
Sample Loader racks.
3.
If a tube has multiple labels on it, spin the tube by hand after it is
put in a rack to be sure it will spin freely and therefore mix
properly. (Refer to
requirements.)
Daily Quality Control Checks
Quality Control checks (which confirm calibration) should be performed
on a daily basis according to the laboratory’s protocol. Commercial
control materials should be properly warmed and mixed according to the
manufacturer’s recommendations. Patient controls should be handled
according to the laboratory’s protocol.
Open Mode QC Procedure
1.
, press [SPECIMEN TYPE].
2.
Move the cursor to the desired QC file and press [QC
SPECIMEN].
3.
If necessary, press [CHANGE SAMPLER] to select the Open
mode.
4.
Run the control.
NOTE: Refer to the
procedures for complete instructions for running samples.
5.
Verify that the results are acceptable.
NOTE: Out of control results are displayed in color.
6.
If the results are unacceptable, repeat the run. If the results are still
unacceptable, obtain a new bottle of the control, be sure that it is
warmed and mixed properly, and again repeat the run. If the results
are still unacceptable, run the other levels of control material. If the
results on all levels are unacceptable, troubleshoot accordingly as
directed in the
7.
When the control results are acceptable, patient samples may be
analyzed.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......