CELL-DYN
3000 System Operator’s Manual
6-7
9140240E — May 1995
Chapter 6
Calibration
WBC, RBC and PLT
Reference values may be determined using multiple counts from a
certified hemocytometer or from a reliably calibrated hematology
analyzer.
Hemoglobin
Reference values may be determined using either the Reference
Cyanmethemoglobin Method or a reliably calibrated hemoglobinometer
or hematology analyzer.
NOTE: DO NOT attempt to calibrate the CELL-DYN 3000
with a hemoglobin standard designed for the calibration of
specific Reference Cyanmethemoglobin Methods. The
instrument uses a Modified Hemiglobincyanide Method which is
not designed to analyze these standards directly.
MCV
Hematocrit Reference Values may be determined from the Reference
Microhematocrit. RBC Reference Values may be determined by using
multiple counts from a certified hemocytometer or from a reliably
calibrated hematology analyzer. The MCV can be calculated from the
Reference HCT and RBC Values.
NOTE: Reference Microhematocrit Values may be determined
by multiple analyses using the NCCLS Method for Packed Cell
Volume (PCV). Use only plain (non-anticoagulated) capillary
tubes. Be certain to verify the proper operation of the
microhematocrit centrifuge and the timer as recommended by
NCCLS (National Committee for Clinical Laboratory
Standards).
Calibration Requirements for Fresh Whole Blood
Minimum requirements for Whole Blood Calibration are described in the
following list.
1.
A minimum of five samples is required for adequate Whole Blood
Calibration.
2.
Samples must be assayed at least in triplicate by reference
methodology and on the CELL-DYN 3000.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......