10-16
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
Figure 10.13:
Vacuum Test Screen
When the [PRESSURE TEST] key is pressed, the system releases the
pressure to atmosphere and then monitors the amount of time required
for it to return to the correct level. The key labels disappear and the
incrementing time is displayed on the screen. When the test is complete,
the time stops incrementing and the key labels are displayed.
The [DIAGNOSTICS] key is used to return to the
When the [INITIALIZATION] key is pressed, the Analyzer is
initialized. This is necessary when a fatal fault has occurred.
When the Analyzer is initialized, a prime cycle must be run.
NOTE: A prime cycle is automatically run whenever the [RUN]
key is pressed after the system is initialized.
DIAGNOSTICS MENU
Vacuum Recovery Time Test
Jan 20 1993
Operator ID
Sequence #
12:39
734
2716
Time elapsed : 3.3
PRESSURE
TEST
DIAGNOSTICS
INITIAL-
IZATION
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......