3-28
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Principles of Operation
Chapter 3
Dispersional Data Alerts
These alerts are triggered by the numeric limits entered into the four
patient limit sets (see the
, for an explanation). If results for a parameter
exceed these limits, they are flagged on the screen and on the report.
Dispersional alerts are displayed or printed as follows:
Screen display:
Result below lower limit shown in yellow
Result above upper limit shown in magenta
Graphic Report:
Results outside limits are underlined
Blank Ticket:
Results outside limits are underlined
Preprinted Ticket:
Results outside limits are marked with an
asterisk
Display and Print
Capacity Exceeded: Result displayed and printed as >>>
Specimens with results that exceed the linearity should be diluted with
Diluent according to the laboratory’s procedure and repeated. (Be sure to
correct the results for the dilution factor used.)
NOTE: MCV, MCH, MCHC and MPV are unaffected by
dilution.
CAUTION: When entering limits into the four patient limit
sets, ensure that the upper and lower limits for each parameter
DO NOT EXCEED the
Linearity Specifications, Chapter 4,
. This will ensure that any value that exceeds these
limits will be flagged for further review.
Suspect Parameter Flags
These flags are generated after the instrument evaluates the measured
data for a particular parameter or group of parameters. The result may be
suspect due to interfering substances or the inability of the instrument to
measure a particular parameter due to a sample abnormality. The name of
each flag, the location of the flag on the display, the cause of the flag and
action to be taken are given in the following explanations.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......