CELL-DYN
3000 System Operator’s Manual
5-11
9140240E — May 1995
Chapter 5 Operating Instructions
When the [REAGENT LOG] key is pressed, one of the logs is displayed
as shown in
. (The name of the displayed log is indicated in the
Status Box). The other two logs may be displayed by pressing the
appropriate soft key. Each log holds 10 entries. The following
information may be entered for each reagent:
Package Size Lot Number
Expiration Date Open Date
Procedure Reagent Log Entry
1.
, press [REAGENT LOG] to display a
2.
Use the soft key to select the appropriate reagent log if it is not
displayed.
3.
Use the Arrow keys on the keyboard to move the cursor to the
desired entry field.
4.
Type the appropriate information.
NOTE: Entries for each field of information are optional. Dates
should be entered with a slash (/) or a period (.) separating the
month, day and year.
5.
Press the Enter key on the keyboard to save the entry and advance
the cursor.
6.
Repeat steps 3–5 until all desired entries have been made.
7.
If desired, press [PRINT LOG] to obtain a printout of the log.
8.
Press the appropriate soft key to select another reagent log and
repeat steps 2–7 to enter data.
Procedure Deleting Entries
When the log is full (the log holds 10 entries), any new entry will replace
the oldest entry. Print the logs when they are full for documentation
purposes.
1.
From the
, press [REAGENT LOG] to display a
2.
Move the cursor to the oldest entry in the log.
3.
Press [DELETE ENTRY]. The [COMPLETE DELETION] and
[RESTORE ENTRY] keys are displayed.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......