10-46
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
Table 10.1:
Instrument Status Conditions
The following abbreviations are used in this table:
BL — Bulletin Line
SB — Status Box
__________________________________________________________
Ready/SB
– – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – -
The Analyzer is ready to process samples.
__________________________________________________________
Selecting open/closed mode/SB
– – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – –
The [CHANGE SAMPLER] key was pressed and the Analyzer is
changing to the selected mode of operation.
Resume processing when the READY message is displayed in the status
box.
__________________________________________________________
Entering standby/SB
– – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – –
The instrument has been idle for four hours and therefore is
automatically performing a cleaning cycle before entering the
STANDBY state. (Pressing the [DAILY SHUTDOWN] key also
initiates this message.)
When the cycle is complete, press [PRIME] or [RUN] to initiate the
priming cycle and return the instrument to the READY state. Resume
operation when the cycle is complete.
__________________________________________________________
Standby/SB
– – – – – – – – –– – – – – Explanation/Action– – – – – – – – – – – – – – -
The instrument has entered the STANDBY state. Press [PRIME] or
[RUN] to initiate the priming cycle and return the instrument to the
READY state. Resume operation when the cycle is complete.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......