6-90
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
•
Analyzer Module Set Points (e.g., gains, dil factors [internal
calibration factors], thresholds, pressure/vacuum settings)
•
RBC Editing Ratio
To back up the calibration factors, proceed as follows:
1.
Turn the Data Station power switch OFF.
2.
Turn the Analyzer power switch OFF.
3.
Obtain the Set-Up Disk from the diskette box located behind the
upper front cover of the Analyzer.
4.
Insert the Set-Up Disk in the Data Station disk drive.
5.
Turn the Data Station power switch ON.
6.
The DATA STATION Screen displays the following:
THIS DISK CAN BE USED TO COPY SET UP
INFORMATION TO OR FROM THE DATA STATION HARD
DISK. THE FOLLOWING COMMANDS ARE AVAILABLE:
GET [ENTER] - COPIES SET-UP INFORMATION FROM
THE DATA STATION HARD DISK ONTO THE SET-UP
DISK
PUT [ENTER] - COPIES SET-UP INFORMATION FROM
THE SET-UP DISK ONTO THE DATA STATION HARD
DISK
7.
Type GET.
8.
Press the Enter key on the keyboard. The Set-Up information is
copied from the Data Station Hard Disk onto the Set-Up Disk.
Previous Set-Up information that was stored on the disk is
overwritten.
9.
When the copy process is complete, the “a:>” prompt is displayed
on the screen.
10.
Remove the Set-Up Disk from the disk drive and return it to the
diskette box.
11.
Turn the Data Station power switch OFF.
12.
Wait five seconds and then turn the Analyzer power switch ON.
13.
Turn the Data Station power switch ON.
14.
When the Data Station Status Box indicates INITIALIZED, press
the [RUN] soft key to Prime.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......