4-8
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
System Specifications
Chapter 4
Operational Specifications — CELL-DYN 3000CS
Operating Environment
Temperature
Patient Samples:
Room Temperature (15–30
o
C)
Monocyte values exhibit a change at lower and
higher temperatures. A 1.5% decrease is seen in
the total monocyte percent at lower temperatures
(<18
o
C). A 6% increase will be seen at higher
temperatures.
Instrument: 15–30
o
C
Background values for RBC and PLT may
exhibit an increase at lower environmental
temperatures (<18
o
C).
Relative Humidity
10% to 85%, RHNC
Complete Cycle Times
Auto-Startup (from STANDBY)
Approximately 5.5 minutes
Auto-Startup (from power OFF)
Approximately 7 minutes*
Run, Open mode
Approximately 40 seconds
(READY to READY)
Shutdown (to STANDBY)
7.5 minutes
*
The laser requires a 15 minute warm up time. If the power has been
OFF longer than 5 minutes, wait 10 minutes after the Auto-Startup
cycle is complete before processing samples.
Approximate Aspiration Volumes (whole blood)
Open mode
170 µL
Closed Sampler
230 µL
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......