CELL-DYN
3000 System Operator’s Manual
10-79
9140240E — May 1995
Chapter 10
Troubleshooting
Table 10.4
Non-Functional Fault Conditions (Continued)
The Sample Loader does not power ON
– – – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
The power cord is loose
1.
Ensure that the power
or disconnected.
cord is securely connected
to the Sample Loader and
verify that it is connected
to the power outlet.
2.
Circuitry malfunction.
2.
Call for technical assistance.
__________________________________________________________
The Sample Loader beeps and the START key is not illuminated
– – – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
The cable that connects
1.
Check that each end of the
the Sample Loader to the
cable is securely connected.
Analyzer is loose or
If necessary, remove the
disconnected.
cable and reconnect it.
Press the INIT key on the
Sample Loader operation
keyboard to initialize the
Sample Loader.
2.
Circuitry malfunction.
2.
Call for technical assistance.
__________________________________________________________
The System does not communicate with LIS
– – – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
Interface malfunction.
1.
2.
System not transmitting.
2.
Call for technical assistance.
or receiving.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......