![background image](http://html1.mh-extra.com/html/abbott/cell-dyn-3000/cell-dyn-3000_operators-manual_450973277.webp)
6-22
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
The [REPEAT SPECIMEN] key (see Figure 6.9) is displayed when the
first run of the first specimen is completed. This key is used to delete all
results (runs) for the current specimen.
NOTE: This key deletes all results for all runs of the current
specimen. It does not delete one run only.
Figure 6.9:
Whole Blood Auto-Cal Results Screen
The [CALIBRATR] key is used to display the
). This screen accesses the Auto-Cal
program that is used to calibrate the instrument with a commercial
calibrator. The following soft key labels are displayed:
EDIT REF VAL
START AUTO-CAL
QUIT AUTO CAL
CLEAR REF VALS
PRINT SUMMARY
RETURN
OPEN
SAMPLER
REPEAT
SPECIMEN
REPEAT
SPECIMEN
INTERRUPT
AUTO-CAL
QUIT
AUTO-CAL
RETURN
WHOLE BLOOD AUTO-CAL
Ready for Calibration
May 11 1992
Operator ID
Sequence #
10:30
734
4100
Open Sampler
RESULTS FOR SPECIMEN #1
Spec 1
Mean
Factor
Parameter
Value
RUN1
RUN2
RUN3
Factor
Factor
%Diff
WBC
7.30
7.42
----
----
0.980
----
----
RBC
4.24
4.39
----
----
0.970
----
----
HGB
12.9
13.0
----
----
0.990
----
----
MCV
90.0
91.0
----
----
0.990
----
----
PLT
244.
254.
----
----
0.960
----
----
SPEC ID CAL#01-01
NO. OF RUNS 3
CALIBRATR
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......