CELL-DYN
3000 System Operator’s Manual
10-49
9140240E — May 1995
Chapter 10
Troubleshooting
Table 10.1
Instrument Status Conditions (Continued)
Westgard Warning — See Levey Jennings/BL.
The message is displayed on the VIEW QC LOG Screen only.
– – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – –
The Westgard Rules were selected during set up of the QC file and the
data in the file has violated one or more of the selected rules.
Review the data in the QC file and take appropriate action.
__________________________________________________________
No Entry Found/BL
– – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – –
The number (sequence number or specimen ID number) entered on the
is not present in the data log.
Check that the entry was correct. If appropriate, enter the correct number.
__________________________________________________________
Duplicated Specimen ID/BL
– – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – –
The work list already contains the specimen ID number that has been
entered.
It is not possible to enter the same specimen ID number twice in the work
list. If appropriate, delete the previous entry and re-enter the number.
__________________________________________________________
Duplicated 4-digit Bar Code ID/BL
– – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – –
The work list already contains the 4-digit bar code number that has been
entered.
It is not possible to enter the same 4-digit bar code number twice in the
work list. If appropriate, delete the previous entry and re-enter the
number.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......