Calibration
Chapter 6
The [START AUTO-CAL] key starts the Auto-Cal program by
initiating three background counts.
The [CONTINUE AUTO-CAL] key is displayed after the background
counts are completed as shown in
NOTE: The key label changes to [INTERRUPT AUTO-CAL]
after the key is pressed.
WHOLE BLOOD AUTO-CAL RESULTS Screen
and the following soft key labels are displayed when the [CONTINUE
AUTO-CAL] key is pressed:
INTERRUPT AUTO-CAL
QUIT AUTO-CAL
RETURN
The [INTERRUPT AUTO-CAL] key is used to interrupt (pause) the
Auto-Cal program.
The [QUIT AUTO-CAL] key is used to exit from the Auto-Cal program
before it is completed. When the [QUIT AUTO-CAL] key is pressed,
the bulletin line displays the message: ALL EXISTING DATA AND
RESULTS WILL BE CLEARED. The following soft key labels are
displayed:
CONFIRM QUIT
CANCEL QUIT
These keys are used to confirm or cancel the Quit Auto-Cal command.
The [PRINT] key is used to print the
The [RETURN] key is used to return to the reference value entry screen.
START
AUTO-CAL
CONTINUE
AUTO-CAL
INTERRUPT
AUTO-CAL
QUIT
AUTO-CAL
RETURN
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......