2.
The cursor is in the <LOT NUMBER> entry field. Type the lot
number and press the Enter key on the keyboard to save the entry
and advance the cursor to the <EXPIRATION DATE> entry field.
3.
This is the same format selected on the
. Separate the digits with a slash (/) or a period (.).
4.
Press the Enter key on the keyboard to save the entry and advance
the cursor to the <WESTGARD RULE SELECTION> entry
fields.
5.
Use the Arrow keys on the keyboard to position the cursor at the
desired Westgard Rule.
6.
Press [TOGGLE ON/OFF] to enable the rule and advance the
cursor.
7.
Repeat steps 5 and 6 until all desired rule selections have been
made.
8.
If desired, press [PRINT] to obtain a printout of the entries.
9.
Press [RETURN] to return to the
Procedure Replicate ID Entry
1.
If necessary, from the
[REPLICATE ID] to display the <REPLICATE ID> entry field.
2.
The cursor is in the <REPLICATE ID> entry field. Type the
Replicate ID number (up to 12 alphanumeric characters may be
entered) and press the Enter key on the keyboard to save the entry
and advance the cursor to the <WESTGARD RULE
SELECTION> entry fields.
3.
Select the Westgard Rules as directed in the
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......