6-74
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
Calibration Factor Calculation
As each run of the whole blood samples is completed, a Calibration
Factor is calculated and updated. This factor is displayed in the column
labeled <SPEC (N) FACTOR>.
The column labeled <MEAN FACTOR> contains the average of the
calibration factors for all of the whole blood samples used in the
calibration. The number in parentheses after mean factor indicates the
number of samples used to compute that mean factor. The <FACTOR %
DIFF> column displays the difference between the calibration factor
currently in the instrument and the mean factor computed by the Auto-
Cal program.
When all runs of all whole blood samples have been completed, the
message:
NEW CAL FACTOR COMPUTED. READY FOR ACCEPTANCE
is displayed on the bulletin line.
1.
Press [PREVIOUS SPECIMEN] and/or [NEXT SPECIMEN] to
review and/or print the results of all whole blood samples analyzed.
NOTE: These results are only available for review and/or
printing before the [ACCEPT MEANS] key is pressed.
2.
Press [PRINT] to obtain a printout of the <SPEC> and <MEAN
FACTORS> and the <FACTOR % DIFF>.
3.
Enter the factor % difference for each parameter in the Mode to
Mode Calibration Criteria Worksheet provided at the end of this
procedure. (This worksheet may be duplicated as needed.) The
NOTE: Delete the sign of the factor % difference value before
entering it in the chart.
4.
To determine which parameters require calibration, continue with
the next
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......