CELL-DYN
3000 System Operator’s Manual
6-27
9140240E — May 1995
Chapter 6
Calibration
Pre-Calibration Procedures
Introduction
It is advisable to perform calibration at a time when it can be completed
without interruption. The Pre-Calibration Procedures in this section
verify proper instrument performance to ensure a successful calibration.
These steps should be completed just prior to beginning the calibration
procedure itself. If problems are detected during these checks, DO NOT
ATTEMPT TO CALIBRATE THE INSTRUMENT. If necessary, CALL
THE CUSTOMER SUPPORT CENTER FOR ASSISTANCE AT 1
(800) CELL DYN. After the problems have been resolved, repeat the
Pre-Calibration Procedures to verify proper performance.
included in this section may be
duplicated as needed.
Calibration Guidelines
1.
date before calibrating the instrument. Instrument cleanliness is
essential for accurate calibration. Therefore, each laboratory should
perform any additional maintenance according to its requirements.
2.
Use only recommended CELL-DYN reagents.
3.
Verify the precision for the Open and Closed modes prior to
calibration as directed in the
Pre-Calibration Procedures Checklist
later in this chapter.
4.
Select and process all whole blood samples according to the
requirements given in the
Whole Blood Calibration Guidelines
section of this chapter.
5.
Be certain that any calibrator vials used are brought to room
temperature and mixed according to the manufacturer’s
instructions given in the package insert.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......