CELL-DYN
3000 System Operator’s Manual
5-71
9140240E — May 1995
Chapter 5 Operating Instructions
•
Results that fall outside the range of the selected limit set are
displayed in color. Yellow indicates that the result exceeded the
lower limit and magenta indicates that the result exceeded the
upper limit. These results are underlined on the graphics and blank
ticket printouts. They are indicated by an asterisk on a preprinted
ticket.
•
Results that exceed the display and print capacity are indicated by
>>>> in place of the result.
•
If an RBC/PLT metering fault occurs, results are suppressed for the
affected parameters and the appropriate CLOG or FLOW
ERROR message is displayed. The upper metering and count
times are also displayed. These messages and times are also printed
in the graphics report.
NOTE: A complete explanation of
Chapter 3, Principles of Operation
•
If a WBC flow error occurs, results are suppressed for the WBC
and Differential and the WBC FLOW error message is displayed.
•
The message SAMPLING ERROR-INCOMPLETE
ASPIRATION is displayed on the Bulletin line if insufficient
sample was detected during aspiration. SAMPLING ERROR is
displayed on the screen and “SAMPLING ERR” is printed on the
graphics report to the right of the MCHC. The same message is
printed to the right of the WBC on the preprinted ticket.
"SAMPLING ERROR" is printed above the list of parameters on
the blank ticket.
NOTE: The Sample Loader automatically stops if four
consecutive Sampling Errors or Metering Faults occur. The
message AUTO SAMPLER CONSECUTIVE DATA FAULT
is displayed on the Bulletin line.
•
The [CLEAR FAULT] key is displayed and a message (e.g.,
DILUENT EMPTY) appears in the Status Box on the Data Station
if a fault condition is detected. The word “Fault” on the Analyzer
status indicator is illuminated in red. The Status Box displays the
message FAULT: SEE DIAG or SEE SPECIAL to direct the
operator to the DIAGNOSTICS or SPECIAL PROTOCOLS
menu for further instructions.
NOTE: After the problem has been corrected, press [CLEAR
FAULT] to resume operation.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......