CELL-DYN
3000 System Operator’s Manual
6-65
9140240E — May 1995
Chapter 6
Calibration
Press [RETURN].
the <COMMENTS> line that instrument calibration was
confirmed and no changes were required. Press the Enter key on
the keyboard to save the entry and advance the cursor.
Be certain to retain all necessary documentation in the instrument
logbook.
2.
Calibration Limit
If the factor % difference for a particular parameter is greater than
the following value, there may be an instrument problem as the
value has exceeded the Calibration Limit.
Results GREATER THAN CALIBRATION LIMIT:
WBC
>10%
RBC
>10%
HGB
>10%
MCV
>10%
PLT
>20%
Check to see if any component that could affect the calibration was
changed as this could cause the factor % difference to exceed the
above limits. For example:
Shear valve
RBC/PLT transducer assembly
WBC flow cell
RBC/PLT aperture
Syringe(s)
Hemoglobin flow cell
If a component was changed, calibrate the instrument as needed
according to the directions in this procedure. If no components
have been changed and the factor % difference exceeds these
limits, DO NOT CALIBRATE; CONFIRM THAT ALL PRE-
CALIBRATION PROCEDURES WERE COMPLETED AND
CALL THE CUSTOMER SUPPORT CENTER FOR
ASSISTANCE AT 1 (800) CELL DYN.
Press [QUIT AUTO-CAL] followed by [CONFIRM QUIT] to
exit Auto-Cal.
Troubleshoot accordingly. Be certain to retain all necessary
documentation in the instrument logbook.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......