CELL-DYN
3000 System Operator’s Manual
5-21
9140240E — May 1995
Chapter 5 Operating Instructions
Figure 5.10: Replicate ID Entry Screen
2.
<LOT NUMBER> and <EXPIRATION DATE> entry fields (see
the following
These entry fields are displayed when the [LOT NUMBER] key is
pressed. This designation is intended for QC Files that are used for
commercial controls.
3.
WESTGARD RULE SELECTION:
RULE 1: Value Outside 3 SD
RULE 2: Two Consecutive values outside SAME 2 SD
RULE 3: Two Consecutive values outside OPPOSITE 2 SD
RULE 4: Two of Three Consecutive values outside SAME 2 SD
RULE 5: Four Consecutive values outside SAME 1 SD
RULE 6: Ten Consecutive values on SAME side of mean.
are discussed in detail in the
The Westgard Rule selections are available on either of the
LOT
NUMBER
RETURN
QC FILE SETUP
Ready
Dec 17 1992
Operator ID
Sequence #
16:36
734
0630
Replicate ID: ------------
FOR Low
WESTGARD RULE SELECTION:
ON
ON
ON
ON
ON
ON
RULE 1:
RULE 2:
RULE 3:
RULE 4:
RULE 5:
RULE 6:
Value outside 3 SD.
Two consecutive values outside SAME 2 SD.
Two consecutive values outside OPPOSITE 2 SD.
Two of three consecutive values outside SAME 2 SD.
Four consecutive values outside SAME 1 SD.
Ten consecutive values on SAME side of mean.
1
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......