5-58
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Operating Instructions
Chapter 5
Data Station Run Screen Soft Keys
A brief description of the function of each key is given in this section.
Instructions for running samples and using the Work List are given in the
Sample Analysis sections of this chapter.
The [CLEAR ORIFICE] key is used to initiate a special cleaning
sequence that flushes the RBC/PLT aperture to remove obstructions. The
sequence takes approximately 35 seconds. When the [CLEAR
ORIFICE] key is pressed, the message CLEANING RBC ORIFICE is
displayed in the Status Box.
The [CLEAR ORIFICE] key label changes to [CLEAR FAULT]
whenever a system fault occurs (e.g., Diluent Empty). This key is used to
clear the fault message and return the Analyzer to the “Ready” status
after corrective action has been taken.
NOTE: A message describing the fault appears in the bulletin
line. A list of fault conditions and corrective action is given in
Figure 5.27:
Work List Screen
CLEAR
ORIFICE
CLEAR
FAULT
WORK LIST
ON
BAR CODE
OFF
INSERT/
DELETE
DELETE
ALL
PURGE
COMPLETED
WORK LIST
SET UP
WORK LIST
RETURN
WORK LIST
Ready
Dec 18 1992
Operator ID
Sequence #
09:48
734
0630
#
793
794
795
796
797
798
799
800
4-DIGIT
SPECIMEN
SPECIMEN
Work List OFF
BAR CODE
ID
NAME
1793
1794
1795
1796
1797
1798
1799
1800
LIMIT PARAM STATUS RACK-TUBE
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Bar Code On
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......