6-52
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
9.
Press [PRINT QC LOG] to obtain a printout of the data.
10.
Press [RETURN] followed by [MAIN] to return to the
Calibration Factor Calculations
1.
Use the mean value for each parameter from the CD MEAN file
printout and the Current Open Mode Calibration Factor printout
from
Preparing for Manual Calibration
calibration factor calculations.
2.
Enter the information from step 1 on the Manual Calibration
worksheet provided at the end of this procedure to calculate the
New Open Mode Calibration Factor for each parameter as follows.
Calibrator Calibration:
Whole Blood Calibration:
3.
Referring to the New Open Mode Cal Factor from the previous step
and the Current Open Mode Cal Factor from the printout, use the
appropriate chart on the worksheet to compute the Factor %
Difference as follows:
New Open Mode Factor - Current Open Mode Factor
x 100
New Open Mode Factor
= Factor % Difference
4.
Enter the factor % difference for each parameter in the Manual
Calibration Criteria chart on the worksheet. The
NOTE: Delete the sign of the factor % difference value before
entering it on the chart.
5.
To determine which parameters require calibration, continue with
the next
Assay Value
CELL-DYN Mean
---------------------------------------------
Current Open Mode
Calibration Factor
×
New Open Mode
Calibration Factor
=
Reference Mean
CELL-DYN Mean
---------------------------------------------
Current Open Mode
Calibration Factor
×
New Open Mode
Calibration Factor
=
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......