CELL-DYN
3000 System Operator’s Manual
12-13
9140240E — May 1995
Chapter 12
Sample Loader
5.
The first sample is then mixed by counter-rotation.
NOTE: Tubes must be able to spin freely while in the tube
racks. Tubes with excessive numbers of labels or with label flaps
sticking out (
) will not spin properly and will trigger
a mixing error. The bulletin line will display the message
SAMPLING ERROR — INCOMPLETE ASPIRATION.
The RUN Screen will display the MIXING ERROR and
SAMPLING ERROR messages.
The Sample Loader does not attempt to aspirate an improperly
mixed sample. Therefore, the SAMPLING ERROR —
INCOMPLETE ASPIRATION message is displayed along with
the MIXING ERROR message and results look like background
counts.
If the tube cannot spin, the Sample Loader emits three beeps, the
mixing head motor stops (to prevent damage to the motor) and the
sample is not aspirated. The message MIXING ERROR is
displayed and printed to the right of the MCH result, and the
message SAMPLING ERROR is displayed and printed to the
right of the MCHC result. An “M” appears in the column preceding
the date in the data log and the QC log.
6.
Rack movement resumes and the first sample is moved to Station 3.
When the aspiration needle begins to move down, the plunger
moves forward to position the tube so the stopper is pierced near
the center. The needle moves down and the Sample Loader signals
the Analyzer to aspirate the sample. The sample is pulled into the
shear valve by action of the sample aspiration pump. The Analyzer
then begins a normal count cycle.
7.
After the sample is aspirated, the aspiration needle is retracted and
washed. The plunger retracts when the needle has been fully
withdrawn from the tube.
8.
-7 are repeated until all tubes in the rack have been
processed.
9.
When the rack being processed reaches the left rear corner of the
tray, the racks are rearranged. The Sample Loader moves the left
front rack over to the right front side. The Sample Loader then
pushes the racks on the right towards the rear and brings the racks
on the left forward. Then the second rack is moved into position on
the right side of the tower and the first tube is positioned at
Station 1.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......