CELL-DYN
3000 System Operator’s Manual
6-3
9140240E — May 1995
Chapter 6
Calibration
General Information
The CELL-DYN 3000 has three modes of operation:
•
Open mode
•
Closed mode — Sample Loader version
•
Closed mode — Closed Sampler version
NOTE: Each CELL-DYN 3000 has only one closed mode of
operation.
Both the Open and Closed modes must be calibrated individually. For
convenience, the Open mode is calibrated first and the Closed mode is
then referenced to it. There are several ways to accomplish the total
calibration, depending on the preference of the user. The following
Calibration Procedural Guidelines section provides a quick reference
guide to the remainder of the chapter.
Calibration Procedural Guidelines
Two methods may be used to calibrate the CELL-DYN 3000:
•
Auto-Cal — automatic calibration program incorporated in the
Data Station software
•
Manual Calibration — an alternative to Auto-Cal
The instrument’s Open mode is calibrated with the calibration material of
choice, using the method of choice. The Closed mode is then calibrated
to correlate with fresh whole blood samples using the Mode to Mode
Calibration method of choice. The
chapter contains an overview for each method of calibration and gives a
detailed explanation of how to use it. Review the appropriate overview
before beginning the selected calibration procedure.
The selected calibration procedure may indicate that the calibration for
some or all of the parameters is acceptable. Therefore, each procedure
includes specific criteria that are used to determine which parameters
require calibration. The criteria are:
•
Validation Range —
Calibration not required
•
Calibration Range —
Calibration is required
•
Calibration Limit —
Do not calibrate, possible instrument
problem exists
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......