5-42
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Operating Instructions Chapter 5
When this option is enabled, scatterplots and histograms are
printed only for samples with flagged results.
4.
Print PCT, PDW results
When this option is enabled, the results for PCT and PDW are
printed on the report.
NOTE: Clinical significance has not been established for these
parameters. Therefore, they are not reportable.
5.
Print X-B Program status
When this option is enabled, the status of the X-B program is
printed on the report.
6.
Print Interpretive Report
When this option is enabled, the Interpretive Report messages are
printed on the report. These messages are generated when results
exceed the Patient Limits and/or instrument-generated flags are
present. Refer to the
Chapter 3, Principles of Operation
7.
Print Limits Report
When this option is enabled, the Patient Limits set that was applied
to the results is printed on the report.
8.
Print Manual Differential Grid for ALERTED specimens
When this option is enabled, a grid that can be used to report a
manual Differential is printed on the report for any specimen that is
flagged.
9.
Print Manual Differential Grid for NON-ALERTED specimens
When this option is enabled, a grid that can be used to report a
manual Differential is printed on the report for any specimen that is
not flagged.
10.
Suppress WBC, RBC, HGB, or PLT value if its flag is set.
When this option is enabled, flagged results are not displayed on
the RUN Screen or in the Data Log.
11.
Line-feeds per page for graphics printer (1 to 99)
This option is used to select the size of the printed report.
(A graphics report typically has 66 lines.)
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......