CELL-DYN
3000 System Operator’s Manual
5-91
9140240E — May 1995
Chapter 5 Operating Instructions
3.
If desired, press [TOGGLE ON/OFF] to select a Specimen Name
entry. The cursor advances to the next entry field.
4.
If desired, press [TOGGLE ON/OFF] to select a Limit Set entry.
The cursor advances to the <DEFAULT LIMIT SET> entry field.
5.
If desired, select a Limit Set to be used as the default (preassigned)
set by typing the number of the desired Limit Set and pressing the
Enter key on the keyboard.
NOTE: If no selection is made, Limit Set 1 will be used as the
default Limit Set.
6.
If desired, press [TOGGLE ON/OFF] to select a Parameter Set
entry. The cursor advances to the <DEFAULT PARAMETER
SET> entry field.
7.
If desired, select a Parameter Set to be used as the default
(preassigned) set by typing the number of the desired Parameter Set
and pressing the Enter key on the keyboard.
NOTE: If no selection is made, Parameter Set 1 will be used as
the default Parameter Set.
Sample Analysis Using the Work List
,
Sample Loader Instruments
NOTE: The Work List is active in the Closed Mode only on the
Sample Loader version of the CELL-DYN 3000.
Using the Work List with Bar Codes
The Sample Loader has a built-in bar code reader that automatically
reads a bar code label placed on the tube. Refer to
complete discussion of bar code labels and instructions for correct
placement of the labels on the tubes.
When the Bar Code ON option is selected, the bar code labeled samples
may be run in any order after the Work List has been created.
When bar code labels are used on the CELL-DYN 3000SL and the Work
List is active, the instrument reads the bar code label and searches the
Work List for that bar code number. When the number is found, the
entered information is displayed on the
stored in the Data Log. If no entry is found, the sample is identified in the
Data Log by “Bar Code Number/B,” as in the following example:
123456/B.
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......