CELL-DYN
3000 System Operator’s Manual
6-51
9140240E — May 1995
Chapter 6
Calibration
Manual Calibration — Open Mode
Procedure
The calibration procedure has been divided into subsections that consist
of a series of easy-to-follow steps. Always follow the entire procedure
unless specifically directed to skip to another section.
NOTE: Be sure to complete the
sections before beginning this procedure.
Preparing for Manual Calibration
1.
, press [CALIBRATION].
2.
If necessary, press [OPEN SAMPLER] to display the Open Mode
Calibration Factors.
3.
Press [PRINT] to obtain a printout of the <CURRENT WHOLE
BLOOD FACTORS> displayed on the screen.
4.
Press [RETURN] followed by [MAIN] to return to the
Determining the Open Mode Mean
1.
, press [RUN].
2.
If necessary, press [CHANGE SAMPLER] to select the Open
mode.
3.
Press [SPECIMEN TYPE].
4.
Use the Arrow keys on the keyboard to move the cursor to an
empty QC File and type “CD MEAN” in the <FILENAME> field.
5.
Press the Enter key on the keyboard to save the name. Use the Up
Arrow key on the keyboard to move the cursor back into the “CD
MEAN” file.
6.
Press [QC SPECIMEN] to return to the
7.
Run each sample the appropriate number of times into the “CD
MEAN” QC file.
NOTE: A minimum of 5 whole blood samples is required.
Whole blood samples must be run at least in triplicate. A
Commercial Calibrator must be run 5—10 times.
8.
With the cursor at "CD MEAN," press [MAIN] followed by [QC
LOGS] and [VIEW QC LOG].
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......