7-16
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Quality Control
Chapter 7
Rule 5 (4
1s
):
Four consecutive values outside same 1SD
Four consecutive results fell outside 1SD on the same
side of the mean
Rule 6 (10x):
Ten consecutive values on the same side of mean
Ten consecutive results fell on the same side of the
mean
The rules may be used singly or in combination depending on operator
preference. Selections are made on the
When a rule is selected, a plus sign is displayed to the right of the
parameter name on the LEVEY-JENNINGS MENU Screens. (See
Figure 7.7.) Six plus signs indicate that all six rules are selected in order
from left to right. A minus sign is displayed if a rule is not selected.
Figure 7.7:
Levey-Jennings Menu Screen Showing Westgard Rule
Violations
The number of the rule that was violated is displayed in place of the plus
sign. Figure 7.7 shows examples of the plus and minus signs and
rule-violation indications.
MPV:1+–+++
10.3
9.3
8.3
PLT:1+–+++
265.
240.
215.
RBC:1+–++6
4.58
4.40
4.22
RDW:1+–++6
17.6
15.4
13.2
LEVEY-JENNINGS MENU
Ready
WBC:1+–++6
8.3
7.7
7.1
GROUP
2
GROUP
3
GROUP
4
NEXT
10
RETURN
Mar 15 1993
Operator ID
Sequence #
10:00
734
2723
QC file: NORM 45 OPEN
Seq num: 9558 to 684
HGB:++–+++
13.8
13.4
13.0
MCV:1+–+++
89.5
86.5
83.5
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......