Chapter 3 Principles of Operation
Action:
Review a stained smear for nucleated red blood cells, platelet
aggregates, sickle cells, or other abnormal cell types and follow
your laboratory's review criteria.
If NRBCs are present, they should be quantitated according to your
laboratory's procedure; correction of the WBC count is not required,
however.
When the NRBC and WBC flags are set concurrently, RBCs resistant to
the lytic action of the sheath reagent may be present. Repeat the sample
using the Resistant RBC specimen type.
RBC Flags
RBC MORPH — displayed next to the MCV result
Cause: One or more of the following parameters exceeds expected
limits:
MCV <80 fL or >100 fL
MCH <25 pg or >34 pg
MCHC <29 g/dL or >37 g/dL
RDW > 18.5%
Action:
Review a stained smear and follow your laboratory's review
criteria.
PLT Flags
No MPV result displayed (data suppressed)
Cause: The PLT histogram did not meet expected criteria (non-log
normal distribution).
Action:
Review a stained smear for abnormal PLT morphology or the
presence of PLT aggregates and follow your laboratory's
review criteria. Verify the PLT count.
Interpretive Messages
Interpretive messages appear only on the graphics report and are
generated when the numeric limits entered in the
, for an explanation). These messages are printed only when
the Interpretive Report option is selected on the
Search
Go Back
Book TOC
Содержание CELL-DYN 3000
Страница 31: ...10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 NOTES Search Go Back Book TOC ...
Страница 265: ...6 10 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 287: ...6 32 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 313: ...6 58 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 347: ...6 92 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Calibration Chapter 6 NOTES Search Go Back Book TOC ...
Страница 351: ...7 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 369: ...7 20 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 371: ...7 22 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Quality Control Chapter 7 NOTES Search Go Back Book TOC ...
Страница 387: ...9 6 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 405: ...9 24 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Maintenance Chapter 9 NOTES Search Go Back Book TOC ...
Страница 431: ...10 2 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 473: ...10 44 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 495: ...10 66 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 509: ...10 80 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Troubleshooting Chapter 10 NOTES Search Go Back Book TOC ...
Страница 535: ...12 16 CELL DYN 3000 System Operator s Manual 9140240E May 1995 Sample Loader Chapter 12 NOTES Search Go Back Book TOC ...
Страница 542: ......