Keysight E5260/E5270 Programming Guide, Edition 4
5- 11
Error Messages
Error Codes
622
Specify trigger output port for OSX.
No trigger output port was specified for the OSX command. Specify the
trigger output port, or set the port as the trigger output port. Refer to
to set trigger port.
630
Incorrect polarity of search step value.
For the linear search measurement. The step value must be positive if
start<stop, or negative if start>stop.
631
Number of search steps must be 1001 or less.
For the linear search measurement. The number of search steps
between start and stop must be 1001 or less. This means the |step| value
must be |stop-start|/1001 or more.
632
Search measurement was aborted.
Search measurement was aborted by the automatic abort function.
640
Search limits must be range/20000 or more.
For the binary search measurement. The limit value for the search
target must be
range
/20000 or more. where
range
means the
measurement range actually used for the measurement.
650
Data format must be ASCII to get time data.
The time stamp function is not available for the binary data output
format. To use the time stamp function, set the data output format to
ASCII.
655
Cannot connect/disconnect series resistor.
The series resistor status cannot be changed in the high voltage state.
Set the output voltage or the voltage compliance to
±
42 V or less to
connect or disconnect the series resistor.
656
Series resistor must be OFF for 1 A range.
The series resistor cannot be set to ON for the measurement channels or
the output channels that use 1 A range.
657
Series resistor cannot be used with ASU.
The series resistor is not available for the channel connected to the Atto
Sense and Switch Unit (ASU).
670
Specified channel does not have ASU.
Specify the module that can be used with the ASU.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......