Keysight E5260/E5270 Programming Guide, Edition 4
4- 37
Command Reference
Command Reference
BSSV
The BSSV command specifies the synchronous voltage source for the binary search
measurement. The synchronous source output will be:
Synchronous source output =
polarity
×
BSV source
offset
where BSV source output means the output set by the BSV command.
This command setting is cleared by the BSI/BSV command.
Execution
Conditions
The BSV command must be sent
before
sending this command.
Syntax
BSSV
chnum
,
polarity
,
offset
[,
Icomp
]
Parameters
chnum
:
Synchronous source channel number. The value must be slot number
where the module has been installed. Integer expression. 1 to 8 (1 and 5
are not available for HPSMU). See
.
polarity
:
Polarity of the BSSV output for the BSV output.
0: Negative. BSSV output =
−
BSV
offset
1: Positive. BSSV output = BSV
offset
offset
:
Offset voltage (in V). Numeric expression.
0 to ±100, or 0 to ±200 for E5280B/E5290A
Both primary and synchronous search sources will use the same output
range. So check the output range set to the BSV command to determine
the synchronous source outputs.
Icomp
:
Current compliance value (in A). Numeric expression. If you do not
specify
Icomp
, the previous value is set. Zero amps (0 A) is not a valid
value for the
Icomp
parameter.
Example
Statements
OUTPUT @E5270;"BSSV 1,0,5,1E-6"
See Also
for the source output value, output range, and the
available compliance values.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......