Keysight E5260/E5270 Programming Guide, Edition 4
4- 149
Command Reference
Command Reference
WZ?
This query command immediately confirms the all channel output, and returns the
status 0 if it is within
±
2 V or 1 if it is more than
±
2 V.
Syntax
WZ? [
timeout
]
Parameters
timeout
:
Timeout. Numeric expression.
0 to 655.35 sec, with 0.01 sec resolution.
With
timeout
parameter, this command waits until the all channel
output becomes within
±
2 V or until the specified
timeout
elapses, and
returns 0 or 1.
The
WZ? 0
command has the same effect as the
WZ?
command.
Query Response
state
<CR/LF^EOI>
0: All channel output is within
±
2 V.
1: Any output channel applies more than
±
2 V.
Example
Statement
OUTPUT @E5270;"WZ? 5.0"
ENTER @E5270;A