Remote Mode Functions
Measurement Modes
Spot Measurements
Spot measurement is performed as shown below. The measurement channel
performs one point measurement.
Figure 2-1
Spot Measurements
1. The source channel starts output by the DV or DI command.
You can use up to eight channels for the 8-ch mainframe, two channels for the
2-ch mainframe.
2. The measurement channel starts measurement by a trigger, such as the XE
command. If the trigger is received during the settling time of the source
channels, measurement starts after the settling time.
You can use up to eight channels for the 8-ch mainframe, two channels for the
2-ch mainframe. If you use multiple measurement channels, the channels
perform measurement in the order defined in the MM command.
3. After measurement, the source channels continue the source output.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
NOTE
The DV command is used to force voltage, and the DI command is used to force
current.
Voltage or current
DV/DI
DV/DI
Measurement trigger (e.g. XE)
Measurement time
Channel 1 output
Channel 2 output
Setup value
Previous value
Setup value
Previous value
Voltage or current
Time
: Measurement
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......