4- 108
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
SSR
This command sets the connection mode of a series resistor (approx. 1 M
Ω
) for each
channel.
If the output switch is opened, the SSR command just sets the mode, and the CN
command connects or disconnects the series resistor.
If the output switch is already closed, the SSR command connects the series resistor
to the SMU output. Then the output forces 0 V one moment.
A series resistor is mounted on each module. If you use a series resistor, the voltage
you set is applied to the near side of the series resistor. Thus, the voltage will be
divided by the series resistor and the device under test.
Execution
Conditions
The series resistor cannot be used for the measurements that use the high resolution
SMU (HRSMU) connected to the Atto Sense and Switch Unit (ASU) or the
measurements that use 1 A range of the high power SMU (HPSMU).
The channel must not be in the HIGH VOLTAGE state (forcing more than ±42 V, or
voltage compliance set to more than ±42 V).
Syntax
SSR
chnum
,
mode
Parameters
chnum
:
Channel number. The value must be slot number where the module has
been installed. Integer expression. 1 to 8 (1 and 5 are not available for
HPSMU). See
.
mode
:
Status of the series resistor. Integer expression.
0: Disconnect (initial setting).
1: Connect.
Example
Statements
OUTPUT @E5270;"SSR 1,1"
OUTPUT @E5270;"SSR 2,1"
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......