3- 12
Keysight E5260/E5270 Programming Guide, Edition 4
Programming Examples
Spot Measurements
Spot Measurements
To perform spot measurements, use the following commands.
NOTE
If you use multiple measurement channels, the channels start measurement in the
order defined in the MM command.
Function
Command
Parameters
Enables channels
CN
[
chnum
... [,
chnum
] ... ]
Disables channels
CL
[
chnum
... [,
chnum
] ... ]
Sets filter ON/OFF
[FL]
mode
[,
chnum
... [,
chnum
] ... ]
Sets series resistor ON/OFF
[SSR]
chnum
[,
mode
]
Sets integration time
(Keysight E5270B can use
AAD/AIT instead of AV.)
[AV]
number
[,
mode
]
[AAD]
chnum
[,
type
]
[AIT]
type
,
mode
[,
N
]
Forces constant voltage
DV, TDV
chnum
,
vrange
,
output
[,
comp
[,
polarity
[,
irange
]]]
Forces constant current
DI, TDI
chnum
,
irange
,
output
[,
comp
[,
polarity
[,
vrange
]]]
Sets voltage measurement
range
[RV]
chnum
,
range
Sets current measurement
range
[RI]
chnum
,
range
[RM]
chnum
,
mode
[,
rate
]
Selects measurement mode
MM
1,
chnum
[,
chnum
... [,
chnum
] ... ]
Sets SMU operation mode
[CMM]
chnum
,
mode
Executes measurement
XE
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......