4- 124
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
TTI
The TTI command performs the high speed spot measurement, and returns the
measurement data and the time data for the time from when the timer is cleared until
the measurement is started. The command starts a current measurement regardless
of the SMU operation mode, trigger mode (TM command), and measurement mode
(MM command).
This command is effective for ASCII data output format. Refer to
.
Execution
Conditions
CN command has been executed for the specified channel.
Syntax
TTI
chnum
[,
range
]
Parameters
chnum
:
Measurement channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
.
range
:
Measurement range or ranging type. Integer expression. If you select
the fixed range, the instrument performs measurement by using the
specified range. For the auto or limited auto ranging, the measurement
range will be set to the minimum range that covers the measured
values. However, the instrument never uses the range less than the
specified range for the limited auto ranging. See
If you do not specify the
range
parameter for voltage output channels,
the channel uses the minimum range that covers the compliance value.
The
range
parameter is meaningless for current output channels. The
measurement ranging type is always same as the output ranging type.
NOTE
To use 1 pA range (only for E5270B)
The measurement channel connected to the ASU (Atto Sense and Switch Unit)
supports the 1 pA range. To use the 1 pA range, set the 1 pA fixed range or the 1 pA
limited auto ranging.
To enable the 1 pA range for the auto ranging mode, execute the
command.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......