Keysight E5260/E5270 Programming Guide, Edition 4
4- 11
Command Reference
Command Parameters
Table 4-3
Current Measurement Ranging Type
range
a
a. If the measurement channel forces current, the channel uses the current output range regardless
of the
range
value.
Ranging type
for measurement mode
without pulse
for measurement mode
that uses pulse
0
Auto ranging
Measurement channel uses
the minimum range that
covers the compliance
value.
8, for E5288A ASU
1 pA limited auto ranging
9, for E5287A
10 pA limited auto ranging
10, for E5287A
100 pA limited auto ranging
11, for E5280B/E5281B/E5287A
1 nA limited auto ranging
12, for E5280B/E5281B/E5287A
10 nA limited auto ranging
13
100 nA limited auto ranging
14
1
μ
A limited auto ranging
15
10
μ
A limited auto ranging
16
100
μ
A limited auto ranging
17
1 mA limited auto ranging
18
10 mA limited auto ranging
19
100 mA limited auto ranging
20, for E5291A
200 mA limited auto ranging
20, for E5280B/E5290A
1 A limited auto ranging
-
8, for E5288A ASU
1 pA range fixed
-
9, for E5287A
10 pA range fixed
-
10, for E5287A
100 pA range fixed
-
11, for E5280B/E5281B/E5287A
1 nA range fixed
-
12, for E5280B/E5281B/E5287A
10 nA range fixed
-
13
100 nA range fixed
-
14
1
μ
A range fixed
-
15
10
μ
A range fixed
-
16
100
μ
A range fixed
-
17
1 mA range fixed
-
18
10 mA range fixed
-
19
100 mA range fixed
-
20, for E5291A
200 mA range fixed
-
20, for E5280B/E5290A
1 A range fixed
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......