Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
WM
The WM command enables or disables the automatic abort function for the staircase
sweep sources and the pulsed sweep source. The automatic abort function stops the
measurement when one of the following conditions occurs:
•
Compliance on the measurement channel
•
Compliance on the non-measurement channel
•
Overflow on the AD converter
•
Oscillation on any channel
This command also sets the post measurement condition for the sweep sources.
After the measurement is normally completed, the staircase sweep sources force the
value specified by the
post
parameter, and the pulsed sweep source forces the pulse
base value.
If the measurement is stopped by the automatic abort function, the staircase sweep
sources force the start value, and the pulsed sweep source forces the pulse base
value after sweep.
Syntax
WM
abort
[,
post
]
Parameters
abort
:
Automatic abort function. Integer expression.
1: Disables the function. Initial setting.
2: Enables the function.
post
:
Source output value after the measurement is normally completed.
Integer expression.
1: Start value. Initial setting.
2: Stop value.
If this parameter is not specified, the sweep sources force the start
value.
Output Data
The E5260/E5270 returns the data measured before any abort condition is detected.
Dummy data 199.999E+99 will be returned for the data after abort.
Example
Statements
OUTPUT @E5270;"WM 2"
OUTPUT @E5270;"WM 2,2"
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......