4- 54
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
ERC
The ERC command changes the output status of the digital I/O port. This command
does not change the status of the trigger ports and the input ports set by the ERM
command.
The *RST command or the device clear sets the digital I/O port (total 16 paths) to
the output port, and sets the port output level to TTL high.
Syntax
ERC
mode
,
value
[,
rule
]
Parameters
mode
:
Control mode. Integer expression. Set
mode
to 2.
2: Controls the digital I/O port.
If you set 1 that is effective for the Keysight 4142B, an error occurs.
value
:
Decimal value of the output status bit pattern. Integer expression. 0 to
65535. The bit pattern must comply with the following rule:
Bit value 0: TTL high level (approx. 2.4 V)
Bit value 1: TTL low level (approx. 0.8 V)
rule
:
Place holder to keep the same syntax as the ERC command of the
Keysight 4142B. Input value is ignored.
Example
Statements
If you want to set TTL low level for the output ports of the digital I/O port bit 0 to 7,
enter the following command.
OUTPUT @E5270;"ERC 2,255"
where the decimal value 255 means binary bit pattern
0000000011111111
. This
command does not change the status of the trigger ports and the input ports.
See Also
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......