Keysight E5260/E5270 Programming Guide, Edition 4
4- 23
Command Reference
Command Reference
AV
This command sets the number of averaging samples of the ADC (A/D converter).
This command is effective for the ADC of the E5260 series and the high-speed
ADC of the E5270B, and not effective for the high-resolution ADC of the E5270A.
This command setting is ignored by the pulsed spot, pulsed sweep, and staircase
sweep with pulsed bias measurements.
Syntax
AV
number
[,
mode
]
Parameters
number
:
1 to 1023, or
−
1 to
−
100. Initial value is 1.
For
positive
number input, this value specifies the number of samples
depended on the
mode
value. See below.
For
negative
number input, this parameter specifies the number of
power line cycles (PLC) for one point measurement. The Keysight
E5260/E5270 gets 128 samples in 1 PLC. Ignore the
mode
parameter.
mode
:
Averaging mode. Integer expression. This parameter is meaningless for
negative
number
.
0: Auto mode (default setting).
Number of samples =
number
×
initial number
1: Manual mode.
Number of samples =
number
where
initial number
means the number of samples the Keysight
E5260/E5270 automatically sets and you cannot change. For voltage
measurement,
initial number
=1. For current measurement, see
If you select the manual mode,
number
must be
initial number
or more
to satisfy the specifications.
Example
Statements
OUTPUT @E5270;"AV 10"
OUTPUT @E5270;"AV -50"
OUTPUT @E5270;"AV 100,1"
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......