Keysight E5260/E5270 Programming Guide, Edition 4
4- 101
Command Reference
Command Reference
RZ
The RZ command returns the channel to the settings that are stored by the DZ
command and clears the stored settings.
The DZ command stores the channel settings (V/I output values, V/I output ranges,
V/I compliance values, and series resistor setting), then sets the channel to 0 V.
Execution
Conditions
The DZ command has been executed for the specified channel. And the CL, CA,
*TST?, *RST or a device clear (HP BASIC CLEAR) command has not been
executed for the specified channel.
Syntax
RZ [
chnum
[,
chnum
...[,
chnum
]...]]
A maximum of eight channels can be set.
Parameters
chnum
:
Channel number. The value must be slot number where the module has
been installed. Integer expression. 1 to 8 (1 and 5 are not available for
HPSMU). See
.
If you do not specify this parameter, this command returns the settings for all
channels that satisfy the conditions described in “Execution Conditions” above, in
the order that the DZ command stored them.
You can specify up to eight channels at once using the RZ command. The
E5260/E5270 returns the stored settings in the order specified.
Example
Statements
OUTPUT @E5270;"RZ"
OUTPUT @E5270;"RZ 1,2,3"
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......