3- 46
Keysight E5260/E5270 Programming Guide, Edition 4
Programming Examples
Multi Channel Sweep Measurements
Multi Channel Sweep Measurements
To perform multi channel sweep measurements, use the following commands.
Function
Command
Parameters
Enables channels
CN
[
chnum
... [,
chnum
] ... ]
Disables channels
CL
[
chnum
... [,
chnum
] ... ]
Sets filter ON/OFF
[FL]
mode
[,
chnum
... [,
chnum
] ... ]
Sets series resistor ON/OFF
[SSR]
chnum
[,
mode
]
Sets integration time
(Keysight E5270B can use
AAD/AIT instead of AV.)
[AV]
number
[,
mode
]
[AAD]
chnum
[,
type
]
[AIT]
type
,
mode
[,
N
]
Sets sweep source timing
parameter
[WT]
hold
,
delay
[,
sdelay
[,
tdelay
[,
mdelay
]]]
Sets auto abort function
[WM]
abort
[,
post
]
Sets voltage sweep source
WV
chnum
,
mode
,
range
,
start
,
stop
,
step
[,
comp
[,
Pcomp
]]
Sets current sweep source
WI
Sets synchronous sweep
source
a
a. The WNX command must be entered after the WV/WI command.
WNX
N
,
chnum
,
mode
,
range
,
start
,
stop
[,
comp
[,
Pcomp
]]
Forces constant voltage
DV, TDV
chnum
,
range
,
output
[,
comp
[,
polarity
[,
crange
]]]
Forces constant current
DI, TDI
Sets voltage measurement
range
[RV]
chnum
,
range
Sets current measurement
range
[RI]
chnum
,
range
[RM]
chnum
,
mode
[,
rate
]
Selects measurement mode
MM
16,
chnum
[,
chnum ...
[,
chnum
]
...
]
Sets SMU operation mode
[CMM]
chnum
,
mode
Executes measurement
XE
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......